How can I evaporate samples with nitrogen without losing labile analytes or contaminating the batch?
Keep the process gentle and controlled. Position the needle so the gas stream plays just above the liquid surface — never submerged — which creates
Keep the process gentle and controlled.
- Position the needle so the gas stream plays just above the liquid surface — never submerged — which creates a soft vortex without splashing.
- Set the bath or block to the lowest temperature that still gives acceptable evaporation speed (typically 30–60 °C for volatile solvent removal) and monitor the endpoint instead of running to complete dryness. Analytes adsorbed on a dry vessel wall are the most common cause of low recovery.
- Use one needle and one position per sample, replace needles regularly and clean the gas manifold between runs to avoid cross-contamination.
For oxygen- or moisture-sensitive compounds, use high-purity nitrogen from your ANVORE generator and seal the vial immediately after concentration.