What vacuum level and temperature should I run for molecular distillation?

Operate with the evaporator temperature high enough to distil the target fraction but below the point where the material decomposes, and keep the condenser surf

Operate with the evaporator temperature high enough to distil the target fraction but below the point where the material decomposes, and keep the condenser surface 40–70 °C cooler than the evaporating surface. Most ANVORE short path systems run in the 0.001–0.1 mbar range; the deeper the vacuum, the lower the required temperature and the lower the thermal stress.

Balance feed rate against wiper speed so the film stays thin and evenly distributed — a feed rate that is too high floods the evaporator and pushes residue into the distillate, while too low a rate starves the film and reduces throughput. Always degas the feed first; dissolved air and light solvents are the most common reason a system will not reach target vacuum.